TRANSIT TIME FLOW METER CASE STUDY
FM500, Sonic-Pro transit time flow meter solves build-up challenge.
transit time flow meters

FM500, transit time flow meter

Application
In a groundwater remediation process at a large oil company located in southern California, the ground water is contaminated by some unknown bacterial/chemical agent.  During the remediation process, the water is pumped through PVC and HDPE pipes, through a series of filters and into a processing chamber where enzymes are injected before being returned to the ground.  To be effective, the enzymes must be accurately injected into the water at a rate which is proportional to the water flow rate.

Problem
Accurate measurements of the water flow rate and total flow is critical to the application.  Initially, paddlewheel type flow meters were installed but a black deposit built up on the paddles, which required cleaning every 2-3 days.  The paddle wheel flow meters were replaced with insertion magnetic flow meters which failed after one day, as the coating proved to be non-conductive.  A turbine flow meter was tried and failed after only a few hours of operation.  A Doppler type ultrasonic flow meter was tested next but it failed to give any measurement at all.

A sample of the water was sent to a test lab for analysis.  The analysis indicated that there were no particles present that could reflect the sound waves making Doppler measurements impossible.  However, Transit-Time test measurements were successful using the FM500, Sonic-Pro hybrid transit time flow meter which can function using either the Doppler or Transit Time measurement methods.  The Sonic-Pro transit time flow meter was also able to measure and display the sound speed of the fluid which, due to the contaminants, was significantly different from plain water.  Knowledge of the actual fluid sound speed is necessary to properly install and calibrate the transit time flow meter.

Besides its tendency to coat the inside wall of the pipe, which over time could possibly block the sound waves, the contaminants also caused a marked reduction of the sound wave intensity in the groundwater, when compared with clean water.  However, the Sonic-Pro's automatic gain function enabled the transit time flow meter to function properly over time by increasing and decreasing the sound wave intensity as needed.

Solution
Clamp-on FM500, Sonic-Pro transit time flow meters were successfully installed at the site on 2 inch diameter PVC and 4 inch diameter HDPE pipes at multiple measurement points.

For another transit time flow meter case study click here.

Click here to learn about the brands in the Analytics family.