U-10 MULTIPARAMETER WATER QUALITY METER
Meter for Monitoring pH, Conductivity, Dissolved Oxygen, Turbidity, Salinity, and Temperature
|

|
FEATURES:
- Rugged
- Fully submersible sensor
- Ideal for multi-parameter water quality measurements without having a permanent installation
|
To Order Call 1-800-876-1172
|
|
Multiparameter Water Quality Meter Product Description
The U-10 multiparameter water quality meter brings laboratory accuracy and the ease of push-button operation to
water quality measurement in the field. Six parameters can be measured : pH, temperature, dissolved oxygen,
electrolytic conductivity, turbidity, and, salinity. Featuring a powerful sensor and state-of-the-art functions, the
multiparameter water quality meter makes it easy to get highly reliable water quality data simply and quickly just by
submersing the sensor in the water. The multiparemeter water quality meter is flexible enough to use for checking
the quality of a broad range of water samples, from factory effluent to urban drainage, river water, lake and marsh water,
aquatic culture tanks, agricultral water supplies, and sea water. The main unit is water resistant, so it can be used
safely even in the rain.
Automatic Calibration
The U-10 multiparameter water quality meter has an automatic calibration function that calibrates the instrument simply
at the touch of a button for more highly reliable data. What used to be a complicated calibration process is now
easy. Just fill the supplied beaker with the standard solution, immerse the sensor and press a button once for one-point
calibration for each of the four parameters pH, conductivity, turbidity and DO. Two-point calibration is also possible
should more precise data be required.
Automatic Salinity Correction
The U-10 multiparameter water quality meter has an automatic salinity function that makes it possible to measure DO in both
fresh water and salt water. It is now possible to measure DO in either fresh or salf water with the same instrument. When
meauring DO in salt water, it will automaticaly measure the salt concentration and make the salinity correction. It is also
possible to set a freely selectable salinity correction factor of between 0 and 4 precent.
Limited Data Logger
20 samples of data for all 6 parameters can be stored in memory to be recalled later for display and print out. Store
the results of measurement of all six parameters for up to 20 samples. There is no need to record measured values on
site. Data stored in memory can be recalled and displayed at any time, and can even be simuItaneously printed out on
any standed printer by connecting it to the Centronics-specification printer output terminals.
The standard handheld multi-parameter water quality measurment system comes with:
- Carrying case with strap for handheld multi-parameter meter and water-quality probe
- Handheld multiparameter water quality meter with water quality probe
- Instruction manual
- pH 4 Buffer
- Reference inner solution
- Calibration beaker
- DO installation tool
Multiparameter Water Quality Meter Specifications
|
PARAMETERS
|
pH
|
CONDUCTIVITY
|
TURBIDITY
|
DISSOLVED OXYGEN
|
TEMPERATURE
|
SALINITY
|
|
PRINCIPLE
|
Glass electrode
|
Alternating four-electrode
|
Scattering/transmitting light
|
Membrane/galvanic cell
|
Thermistor
|
(Alternating four-electrode)*
|
|
RANGE
|
0-14 pH
|
0-100 mS/cm
|
0-800 NTU
|
0-19.9 mg/l
|
0-50° C
|
0-4 %
|
|
RESOLUTION
|
STANDARD
|
0.1 pH
|
(mS/cm) (mS/cm)
0-1 :0.01
1-10 :0.1
10-100 :1
|
10 NTU
|
0.1 mg/l
|
1° C
|
0.1%
|
|
EXPANDED
|
0.01 pH
|
(mS/cm) (mS/cm)
0-1 :0.001
1-10 :0.01
10-100 :0.1
|
1 NTU
|
0.01 mg/l
|
0.1° C
|
0.01%
|
|
REPEATABILITY
|
+0.05 pH
|
+1% of full scale
|
+3% of full scale
|
+0.1 mg/l
|
+0.3° C
|
+0.1 %
|
|
TEMP. COMPENSATION
|
0-50° C
|
0-50° C
|
N/A
|
0-40° C
|
N/A
|
N/A
|
|
CALIBRATION
|
Auto one-point calibration, manual two-point calibration.
|
Auto one-point calibration, manual two-point calibration.
|
Auto one-point calibration, manual two-point calibration.
|
Auto one-point calibration(air), manual two-point calibration.
|
N/A
|
N/A
|
|
DISPLAY
|
LCD
|
|
PRINTER OUTPUT
|
Centronics
|
|
POWER
|
Battery 6F22(S-006P)x1pc.
|
|
SENSOR ASSEMBLY
|
Standard: sensor with 2 m cable. Optional: Sensor with 10 m or 30m cable.
|
|
WEIGHT
|
Main Unit: ~400g(0.9lbs), Sensor Assembly (2 m cable): ~800g(1.8lbs)
|
|
Note:
* Salinity is calculated from the conductivity data
Multiparameter Water Quality Meter Options & Prices
U-10-2 Multiparameter Probe
Probe includes 2 m cable. Note: Cable is not detachable.
|
$2,850 |
|
|
U-10-10 Multiparameter Probe (with 10 m cable)
|
$3,200 |
|
|
U-10-30 Multiparameter Probe (with 30 m cable)
|
$4,100 |
|
|
U-10 REPLACEMENT PARTS
|
|
pH sensor
|
$86 |
|
|
DO sensor
|
$369 |
|
|
Reference sensor
|
$80 |
|
|
RELATED PRODUCTS
|
pH/Ion 3400i Ion Selective Electrode Analyzer
Electrodes not included.
|
$1,238.66 |
|
U-22XD-2 Multiparameter Probe
Use the multiparameter probe to monitor 10 different parameters, including total suspended solids, water depth, and ORP.
|
$5,320 |
|
W-23XD-2 Water Quality Sonde
Water quality instrumentation for simultaneously measuring up to 13 different parameters.
|
$7,270 |
|
|
Water quality sensors
for permanent installation.
|
$Varies |
|
|
|